| Catalog Number |
305062 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The date received by manufacturer has been used for this field.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that bd¿ luer-lok¿ blunt fill needle had an oily substance on the plunger.This was discovered when the medication turned brown after being drawn into the syringe.The medication was propofol , which is white in color.There was no report of injury or further medical intervention.
|
| |
|
Event or Problem Description
|
|
It was reported that bd¿ luer-lok¿ blunt fill needle had an oily substance on the plunger.This was discovered when the medication turned brown after being drawn into the syringe.The medication was propofol , which is white in color.There was no report of injury or further medical intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Since no samples displaying the condition reported are available for examination, we were unable to fully investigate this incident, therefore a root cause could not be determined.However, the customer is likely describing silicone presence on the stopper, which is normal and expected.Furthermore, a device history record review showed no rejected inspections or quality issues during the production of the provided lot number that could have contributed to the reported defect.We would be very interested in examining product that does not meet your expectations and our quality standards.Should you again experience any problems with our product we would appreciate the opportunity to conduct a thorough analysis.Examination of the product involved may provide clarification as to the cause for the reported failure.Please note that silicone is an inert, non-toxic medical substance used as a lubricant for disposable hypodermic products.It is an integral part of the syringe, enabling it to perform as required in various clinical applications and does not present a safety or efficacy issue nor does it impact product function.The silicone application process is designed to provide an even distribution of silicone on the interior of the syringe barrel.Silicone has been in use in this application for over 20 years.No reports are known of adverse clinical effects associated with these products and unintentional delivery of silicone fluid lubricant we appreciate you taking the time to bring this observation to our attention.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to an it issue beginning on (b)(6) 2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other.Device single use?: no.Device returned to manufacturer: no.
|
| |
|
Event or Problem Description
|
|
It was reported that bd luer-lok blunt fill needle had an oily substance on the plunger.This was discovered when the medication turned brown after being drawn into the syringe.The medication was propofol, which is white in color.There was no report of injury or further medical intervention.
|
| |
|
Search Alerts/Recalls
|