Model Number METRIX AIR
Medical Device Problem Code Low Test Results (2458)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report # (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause: mlc-58 - user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Test strip (b)(4).Note: manufacturer contacted customer on 7/19/2018 in a follow-up call to ensure that the replacement products resolved the initial concern; customer stated that the replacement product is working to their satisfaction and they have not had any medical intervention or symptoms since the last call.
 
Event or Problem Description
Consumer reported complaint for low blood glucose test results.The customer is concerned with test results from results obtained of 54, 65 and 83 mg/dl.The customer's expected fasting blood glucose test result range is 90 - 104 mg/dl.The customer feels well and did not report any symptoms.Customer stated when she obtained the low results she did not have any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call on (b)(6) 2018, a back to back blood test performed by the customer non-fasting and produced test results of 105 mg/dl and 113 mg/dl using truemetrix air meter.The product is stored according to specification in the bedroom.The test strip lot manufacturer's expiration date is 09/23/2019 and open vial date is two weeks.The meter memory was reviewed for previous test result history: (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
METRIX AIR
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7748566
Report Number1000113657-2018-00828
Device Sequence Number1084411
Product Code NBW
UDI-Device Identifier00041415072787
UDI-Public(01)00041415072787
Combination Product (Y/N)N
Initial Reporter StateFL
PMA/510(K) Number
K150052
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Replace
Type of Report Initial
Report Date (Section B) 08/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/23/2019
Device Model NumberMETRIX AIR
Device Catalogue NumberREA4H01-01
Device Lot NumberMV2831
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2018
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2018
Initial Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-