| Model Number |
MZ9271 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/22/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The affected product has been received and the evaluation is pending.A follow up report will be submitted once the evaluation is completed.
|
| |
|
Event or Problem Description
|
|
The customer reported fluid leaked at uac junction.The uac was discontinued.There was no report of patient harm.
|
| |
|
Event or Problem Description
|
|
The customer reported fluid leaked at uac junction.The uac was discontinued.There was no report of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer¿s report that the tubing fluid leaked at the uac junction was confirmed.Functional pressure testing was able to confirm a leak from the suspect maxzero female port.Closer inspection of the maxzero under magnification observed a microchannel on the interior maxzero walls that allowed fluid to leak.The cause of the leak is a microchannel found in the maxzero component.The root cause of the microchannel which creates a fluid path is a manufacturing issue.
|
| |
|
Event or Problem Description
|
|
The customer reported fluid leaked at uac junction.The uac was discontinued.There was no report of patient harm.
|
| |
|