Catalog Number PNS5F130BER
Medical Device Problem Codes Device Damaged Prior to Use (2284); Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/03/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the 5f select was ovalized approximately 93.0 thru 99.5 cm from the hub.Conclusions: evaluation of the returned 5f select revealed that the catheter shaft was ovalized.If the device is forcefully retracted out of its packaging at extreme angles or otherwise mishandled during preparation, damage such as this may occur.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
During preparation for a medical procedure, the physician found that the neuron select catheter 5f (5f select) was flattened in the extremity upon removal from the packaging.The damage to the 5f select was found prior to use.Therefore, the 5f select was not used in the procedure.The procedure was completed using a new 5f select.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEURON SELECT CATHETER 5F
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7749488
Report Number3005168196-2018-01518
Device Sequence Number16570457
Product Code DQY
UDI-Device Identifier00814548010212
UDI-Public00814548010212
Combination Product (Y/N)Y
Initial Reporter CountryFR
PMA/510(K) Number
K083125
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/12/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/21/2021
Device Catalogue NumberPNS5F130BER
Device Lot NumberF82488
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/12/2018
Initial Report FDA Received Date08/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-