• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER; MAGIC 3 CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 MAGIC3¿ INTERMITTENT CATHETER; MAGIC 3 CATHETER Back to Search Results
Model Number 53516
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Per direction by fda in a letter dated, july 20, 2018, this emdr is being filed to submit new information obtained based on the complaint investigation results.The initial report for this event was submitted in the quarter 1 2018 asr report.However, supplemental reports regarding the asr are no longer being accepted.It was agreed upon with the fda that the date of awareness would be the date of the letter received, which is july 20, 2018.Received 5 magic3 intermittent catheter lot samples and 1 comparable sample for evaluation.Per visual inspection, samples were visually inspected for holes or damage on the package, legibility of printed numbers and letters, expiration date, correct impression, print color, missing parts of printing, position of printing, damaged component, foreign matter.The reported complaint was confirmed since 1 of the 5 evaluated samples failed the lubricity test.Lubricity test was performed for all of the samples.The reported complaint was confirmed as manufacturer related.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the catheter lacked lubricity.The patient reported that he was unable to use the catheters without applying additional lubricant.Reason for report: investigation results.Report id number: (b)(6).Initial asr report submitted: quarter 1/2018.Corresponding exemption number: e1997042.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGIC3¿ INTERMITTENT CATHETER
Common Device Name
MAGIC 3 CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7750732
Report Number1018233-2018-03339
Device Sequence Number19258317
Product Code MJC
UDI-Device Identifier00801741068522
UDI-Public(01)00801741068522
Combination Product (Y/N)N
Initial Reporter StateUT
Initial Reporter CountryUS
PMA/510(K) Number
K033477
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 08/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/28/2022
Device Model Number53516
Device Catalogue Number53516
Device Lot NumberJUBW0376
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/04/2018
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/20/2018
Initial Report FDA Received Date08/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-