| Model Number |
MMT-522RNAS |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/08/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Currently it is unknown whether or not the device may have caused or contributed to the event.The device has been returned, but not yet evaluated.Further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Customer reported via phone call that the insulin pump alarmed compromised force sensor system.The customer¿s blood glucose level was 126 mg/dl.Customer also reported that the insulin was squirting out during manual prime process.Customer advised during the seating of the force the sensor did not detect the reservoir.The customer was advised that the insulin pump needed to be replaced and was advised to discontinue use of the insulin pump and to revert to the backup plan.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received compromised force sensor system alarm during the basic occlusion test due to loose or protruded drive support disk.Unable to perform prime test, excessive no delivery test and occlusion test due to compromised force sensor system alarm.
|
| |
|