MAUDE Adverse Event Report: JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON AND JOHNSON SURGICAL VISION, INC. TECNIS ITEC PRELOADED 1-PIECE IOL; MONOFOCAL IOLS Back to Search Results
Model Number PCB00
Medical Device Problem Codes Difficult or Delayed Positioning (1157); Malposition of Device (2616)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/20/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If explanted, give date: not applicable as to date the lens remains implanted and therefore not explanted.(b)(6).All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was prepared perfectly.During implantation, the lens was inserter into the patient's operative eye but the trailing haptic got stuck in the cartridge.With the mcpherson pincet the haptic was released but in the meantime the lens was turned.Reportedly, the surgeon managed to turn over the lens in the eye.The lens remains implanted.No patient injury no wound enlargement was reported.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 08/25/2018.Device returned to manufacturer? yes.Device evaluation: the preloaded delivery system , model pcb00, was returned at the manufacturing site for evaluation.Visual inspection under the microscope showed scarce amount of viscoelastic residue on the pcb00 device.The dfu (directions for use) states to completely fill the viewing window of the pcb00 with ovd (ophthalmic viscosurgical device).The intraocular lens (iol) was not returned.The pcb00 device was opened and no assembly defects were observed.There were no damages observed in the assembly.The customer's reported complaint was not verified.No product quality deficiency was identified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search revealed that one similar complaint for this order number have been received.However, there was no associated deviation or non-conformity report found related to this complaint.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON AND JOHNSON SURGICAL VISION, INC.
santa ana CA
MDR Report Key7751068
Report Number2648035-2018-01088
Device Sequence Number1835872
Product Code HQL
UDI-Device Identifier05050474558182
UDI-Public(01)05050474558182(17)210418
Combination Product (Y/N)Y
Initial Reporter CountryNL
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/18/2021
Device Model NumberPCB00
Device Catalogue NumberPCB0000160
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/25/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer09/24/2018
10/25/2020
Initial Report FDA Received Date08/06/2018
Supplement Report FDA Received Date10/22/2018
11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-