| Model Number |
179704990 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/05/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination found that the threads on the poly screw¿s tulip head had become torn off.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the tulip head threads becoming torn off cannot be positively determined.However, noted damage suggests that inadvertently cross threading occurred.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Screw was inserted with correct screw driver and no issues.Rod was placed in construct and locking caps were inserted.The locking caps for the sai screw could not be tightened although rod seemed to be seating well in screw head.A second locking cap was taken trying again to lock it.The it was realized that the issue was with the screw and not the locking cap.Accordingly the screw was replaced with a new one and everything could be completed without a problem.Patient consequence?: no.Action taken for procedure:new screw was inserted.Is the information being submitted for this complaint all the details that are known/available regarding this event?: yes.
|
| |
|
Search Alerts/Recalls
|