| Catalog Number |
232014 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the suture was broken when the tension was applied in the bone hole during acl reconstruction surgery on (b)(6) 2018.The surgery was finished without any other problem although it was not reported how the surgery was completed.It was brand new and the first use when the issue occurred.There was less than 30-mins.Surgical delay and no harm to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Evaluation statement: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.A non-conformance search was performed for this product code 232014 lot l458482 combination and no non-conformances were identified.No further information regarding the technique or instruments used has been provided to determine a root cause for this failure.If any additional information is obtained, this complaint will be re-opened to capture that information.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the suture was broken when the tension was applied in the bone hole during acl reconstruction surgery on (b)(6) 2018.The surgery was finished without any other problem although it was not reported how the surgery was completed.It was brand new and the first use when the issue occurred.There was less than 30min.Surgical delay and no harm to the patient.Additional information from affiliate 8/7/18: suture color?: no further information was provided.Where on suture did the break/damage occur?: at contacting part with tendon.Was suture in contact with an instrument during breakage?: no further information was provided.Open or arthroscopic procedure? no further information was provided.Was a suture passer used in the procedure? what one?: no further information was provided.What suture cutter was used? no further information was provided.Did the surgeon use proper suture management?: no further information was provided.How was the procedure completed?: no further information was provided.
|
| |
|