| Medical Device Problem Code |
Pumping Problem (3016)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The pump has not been returned to animas for evaluation.If the device is returned, an evaluation shall be completed and a supplemental report will be filed.No conclusions can be made at this time.
|
| |
|
Event or Problem Description
|
|
On (b)(6) 2018, the reporter contacted animas, alleging a history/settings (inaccurate delivery) issue.The reporter alleged the basal rates being delivered were not the same as what was programmed.The reporter alleged she did not change her basal rates.There was no indication that the product caused or contributed to an adverse event.This issue is being reported as an issue with the pump not performing as intended with regards to the history or the settings may result in over or under delivery.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the pump has been returned and evaluated by product analysis on 01-may-2019 with the following findings: during investigation, a review of the total daily dose history showed incongruent dates, changing from (b)(6)2019 to (b)(6)2018.There was no data in the black box from these dates due to continued use of the pump.The daily insulin delivery totals appeared inconsistent due to the date changes.The pump history showed the pump was delivering programmed the basal rate until deliveries were halted by the user at 11:32 pm on (b)(6)2018.During testing, the pump successfully completed a rewind, load, and prime sequence.The pump passed the delivery accuracy testing and was found to be delivering within required specifications.All activity performed during testing was accurately recorded in pump history.The complaint of a history settings issue was not duplicated during investigation.There was no defect found.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.
|
| |
|
Search Alerts/Recalls
|