Model Number 383401
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/19/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer reported the feeding pump had a blank screen.When their biomed department went to check the device, the unit would not power on.The event was discovered during testing.There is no reported patient involvement.After follow-up request for more information, the customer states all that was written from biomed was "unable to read the screen".The screen seemed intact, without defect but nothing was displayed on the lcd.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
An evaluation of the kangaroo pump was performed for a blank screen.Liquid ingress caused corrosion, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Common Device Name
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
2824 airwest blvd
plainfield IN 46168
MDR Report Key7753941
Report Number3008361498-2018-00407
Device Sequence Number16200306
Product Code LZH
UDI-Device Identifier10884521144019
UDI-Public10884521144019
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/26/2018
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer08/08/2018
08/08/2018
Initial Report FDA Received Date08/06/2018
Supplement Report FDA Received Date10/02/2018
10/12/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-