Model Number TRUEMETRIX
Medical Device Problem Codes Unable to Obtain Readings (1516); High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/11/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal report # (b)(4).Product not returned for evaluation.No replacement product needed at this time.Complaint resolved by training during the time of the call.Product is working as designed.Most likely underlying root cause: mlc-18 - user has high glucose value.Test strip (b)(4).Note: at follow-up call was made on 07/13/2018; customer stated that she did go to the hospital and that she did not want to talk about it.Customer disconnected the call.Unable to verify the reason that the customer reported visiting the hospital.Unable to contact the customer via telephone at call backs on 07/16/2018 and 07/17/2018.Unable to send product notification letter as no address on file.
 
Event or Problem Description
Consumer reported complaint for e-5 error code.The e-5 error occurred when the blood sample was absorbed.During the call on (b)(6) 2018, a blood test was performed by the customer and produced test result of hi using truemetrix meter.The customer feels well and did not report any symptoms.Customer was going to go the hospital and was not able to continue the troubleshooting process.No further information was able to be obtained.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUEMETRIX
Common Device Name
BLOOD GLUCOSE SYSTEM
Manufacturer (Section D)
TRIVIDIA HEALTH, INC.
2400 nw 55th court
fort lauderdale FL 33309
MDR Report Key7754050
Report Number1000113657-2018-00833
Device Sequence Number16574174
Product Code NBW
UDI-Device Identifier00052569138557
UDI-Public(01)00052569138557
Combination Product (Y/N)N
PMA/510(K) Number
K140100
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date (Section B) 08/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date09/28/2019
Device Model NumberTRUEMETRIX
Device Catalogue NumberRE4H01-81
Device Lot NumberMV2836
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Home
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/11/2018
Initial Report FDA Received Date08/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-