• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Catalog Number 400SMTXSFT2H04
Medical Device Problem Codes Failure to Advance (2524); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the pusher assembly.The pusher assembly was kinked approximately 33.5, 35.0, 44.0, and 45.5 cm and bent approximately 5.5, 12.0, 58.0, 79.0, 118.5, and 127.0 cm from the proximal end of the pusher assembly.The embolization coil was stretched but intact with the pusher assembly.Dried coagulated blood was found along the length of the embolization coil and pusher assembly.Conclusions: evaluation of the returned smart coil revealed a stretched embolization coil.Blood was found coagulated on the embolization coil and pusher assembly.If blood coagulated on the device, it may have contributed to resistance and the reported inability to advance the device into the microcatheter.If the device is forcefully advanced and retracted while experiencing resistance, damage such as stretching of the embolization coil may occur.Further evaluation revealed kinks and bends on the pusher assembly.If the device is advanced while experiencing resistance, damage such as kinks and bends may occur.Based on the return condition, some of the kinks and bends are likely incidental to the complaint and may have occurred during packaging for return to penumbra.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure using penumbra smart coils (smart coils).During the procedure, the physician successfully placed multiple smart coils using a non-penumbra microcatheter.The physician experienced resistance and then was unable to advance a smart coil through the hub of the microcatheter and, therefore, the smart coil was removed.The procedure was completed using a new smart coil and the same microcatheter.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key7754303
Report Number3005168196-2018-01525
Device Sequence Number13872812
Product Code HCG
UDI-Device Identifier00814548015323
UDI-Public00814548015323
Combination Product (Y/N)Y
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/09/2018
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2022
Device Catalogue Number400SMTXSFT2H04
Device Lot NumberF80833
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 07/09/2018
Initial Report FDA Received Date08/06/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age55 YR
Date Report Sent to FDA01/01/2005
-
-