Model Number H7493912420250
Device Problems Material Puncture/Hole (1504); Material Rupture (1546)
Patient Problem No Patient Involvement (2645)
Event Date 07/17/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.  (b)(4).
 
Event Description
It was reported that catheter perforation and balloon rupture occurred.The target lesion was located in the right coronary artery.A 20mm x 2.50mm nc quantum apex¿ balloon catheter was selected for dilatation.However, the guide wire exited the balloon side instead of going through the monorail segment and it also ruptured the balloon.The device was removed and the procedure was completed with a different device.No patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX¿
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key7754361
MDR Text Key116154697
Report Number2134265-2018-07084
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/05/2020
Device Model NumberH7493912420250
Device Catalogue Number39124-2025
Device Lot Number21225573
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-