| Catalog Number |
112230 |
| Medical Device Problem Codes |
Detachment of Device or Device Component (2907); Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/26/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Broken screw pelvic array.The set is heavily used and the surgeon uses a lot of force tightening the screw.Another pelvic array was available to complete the case.
|
| |
|
Event or Problem Description
|
|
Broken screw pelvic array.The set is heavily used and the surgeon uses a lot of force tightening the screw.Another pelvic array was available to complete the case.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: broken screw pelvic array.The set is heavily used and the surgeon uses a lot of force tightening the screw.Another pelvic array was available to complete the case product evaluation and results: visual inspection.Visual inspection shows thumb screw damage.See attachment 1 for image of the instrument.Dimensional inspection: dimensional inspection was not completed since visual and functional inspection confirmed damaged thumb screw.Functional inspection: functional inspection confirmed pelvic array assembly, p/n 112230 damaged thumb screw not engaging properly.Product history review: review of the device history records indicate (b)(4) were manufactured under lot 19470117 and accepted into final stock on 04/07/2017.No non-conformances were identified during inspection.Complaint history review: a review of complaints in catsweb and trackwise related to p/n 112230, lot number 19470117 shows no additional complaints related to the failure in this investigation.Conclusions: visual inspection confirmed damaged thumb screw.Functional inspection confirmed thumb screw not engaging onto array adapter.Corrective action/preventive action: no action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.
|
| |
|