| Catalog Number |
03L79-31 |
| Medical Device Problem Code |
High Readings (2459)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.There is no further patient information provided by the customer.
|
| |
|
Event or Problem Description
|
|
The customer reported falsely elevated calcium results on two patients generated on the architect analyzer.The results provided were: pt#1 on 13jul2018 ca =14.5 / repeated x2 = 10.2 / 10.0mg/dl (normal range = 7.9-10.7mg/dl).There was no reported impact to patient management.
|
| |
|
Additional Manufacturer Narrative
|
|
The calcium assay was removed as a cause of the customer's issue on 08august2018.The architect c4000 analyzer remains as the potential cause reported under mdr # 1628664-2018-01270 and all further information will be documented under that mdr number.Based upon this new information, this complaint is no longer a reportable event.
|
| |
|