| Catalog Number |
48144530 |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Codes |
Neurological Deficit/Dysfunction (1982); Injury (2348)
|
| Date of Event |
07/12/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device was not returned to manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was removing screws and the patient lost motor and ssep monitoring.A screw appeared to be cold welded with the tulip stuck to the screw shank in an angled position.The surgeon was able to remove the tulip and screw from the patient.The case has been aborted.Upon follow up, it was reported that the patient had a neurological deficit on one leg post operatively.
|
| |
|
Additional Manufacturer Narrative
|
|
Product inspection could not be performed as the reported screw was discarded by the hospital staff and not available for evaluation.No lot # was provided, so a manufacturing record review could not be performed.No x-ray images were provided that would help clarify if the screw went too deep or wrong trajectory while removing it.Due to lack of information provided, the exact root cause of the reported event cannot be determined.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was removing screws and the patient lost motor and ssep monitoring.A screw appeared to be coldwelded with the tulip stuck to the screw shank in an angled position.The surgeon was able to remove the tulip and screw from the patient.The case has been aborted.Upon follow up, it was reported that the patient had a neurological deficit on one leg post operatively.
|
| |
|