• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND ARCHITECT AFP; ALPHA-FETOPROTEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ARCHITECT AFP; ALPHA-FETOPROTEIN Back to Search Results
Catalog Number 03P36-25
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event or Problem Description
The customer observed falsely elevated alpha-fetoprotein results while using architect afp reagents.The following data was provided.Sid (b)(6) initial 9.96 ng/ml, repeat 10.15 ng/ml.Previously, the patient results ranged from 200 to 500 ng/ml.No change to surgery or medication was reported due to the elevated values.No impact to patient management was reported.
 
Additional Manufacturer Narrative
Evaluation of the customer issue included a complaint text review, a search for similar complaints, device history review, instrument log review, labeling review, field data review, and accuracy testing.Returns were not available.No adverse trend was identified for the customer issue.Device history review did not identify any issues that may have caused the customer issue.Log review did not identify any issues that contributed to the customer issue.Labeling was reviewed and found to be adequate.Abbott link world wide data was used to evaluate historical product field performance.Patient data was analyzed and compared to an established control limit.The evaluation indicated that the patient median result for the complaint lot was within the established control limits and performance was comparable to other lots in the field with other lots in the field between february 18, 2018 and august 18, 2018.No unusual reagent lot performance was identified.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.Based on all available information and abbott diagnostics evaluation, no systemic issue was identified and no product deficiency was found.Section b5 was corrected.The description of the issue, was corrected as the customer alleged falsely depressed results and not falsely elevated results.The values did not change.
 
Event or Problem Description
Updated september 5, 2018.The customer observed falsely depressed results and not falsely elevated results.The values reported did not change.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT AFP
Common Device Name
ALPHA-FETOPROTEIN
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
MDR Report Key7754528
Report Number3008344661-2018-00068
Device Sequence Number1938553
Product Code LOK
UDI-Device Identifier00380740081317
UDI-Public00380740081317
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120008
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/24/2018
Device Catalogue Number03P36-25
Device Lot Number81397FN00
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer09/05/2018
Initial Report FDA Received Date08/06/2018
Supplement Report FDA Received Date09/05/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARCHITECT I1000SR ANALYZER; ARCHITECT I1000SR ANALYZER; LN 01L86-01 SN I1SR52994; LN 01L86-01 SN I1SR52994
-
-