| Catalog Number |
0250070443 |
| Health Effect - Clinical Code |
Burn, Thermal (2530)
|
| Date of Event |
07/02/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device manufacture date is not known at this time.However, should it become available it will be provided in future reports.Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a burn to the patient.Please note the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.Alleged failure: inner sheath and l-tip started to melt during a procedure.The surgeon cauterized the liver when it was not meant to be cauterized.The failure(s) identified in the investigation is consistent with the complaint record.The probable root causes could be, conductive irrigating or aspirating fluid used was inside the distal end, and contacted the probe and the sheath causing the sheath to melt, power set too high.The failure mode will be monitored for future reoccurrence.
|
| |
|
Event or Problem Description
|
|
It was reported that there was a burn to the patient.Please note the procedure was completed successfully.
|
| |
|