• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: KARL STORZ SE & CO. KG THERMOFLATOR; INSUFFLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

KARL STORZ SE & CO. KG THERMOFLATOR; INSUFFLATOR Back to Search Results
Model Number 26432020-1
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Low Blood Pressure/ Hypotension (1914)
Event Date 06/06/2018
Event Type  Injury  
Manufacturer Narrative
The thermoflator unit was evaluated.We found the unit functions properly and the pressure and flow are within specifications.
 
Event Description
Allegedly, during a laparoscopic procedure, while the patient's abdomen was being insufflated, the thermoflator unit's set pressure was not reading, and the patient became hypotensive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOFLATOR
Type of Device
INSUFFLATOR
Manufacturer (Section D)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM 
Manufacturer (Section G)
KARL STORZ SE & CO. KG
dr.-karl-storz-strasse 34
78532
tuttlingen,
GM  
Manufacturer Contact
susie chen
2151 e. grand avenue
el segundo, CA 90245-5017
4242188201
MDR Report Key7754813
MDR Text Key116208109
Report Number9610617-2018-00060
Device Sequence Number1
Product Code HIF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K955073
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 06/07/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number26432020-1
Device Catalogue Number26432020-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/25/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/19/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-