| Model Number |
FC1SP |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
During evaluation, it was noted that broken fibers were present throughout the cable.Additionally, the cable exhibited a concentrated number of broken fibers which indicates mishandling of the cable.The concentration of broken fibers led to an elevated temperature.Invuity's ifu clearly states to handle the cables with care and not to coil the cables tighter than six (6) inches.
|
| |
|
Event or Problem Description
|
|
During a hip replacement procedure, it was observed that the drape had a burn mark where the fiber optic cable had been resting upon it.There was no injury to the patient.
|
| |
|