| Catalog Number |
RLT281218 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
07/16/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2018, the patient presented with an abdominal aortic aneurysm that was intended to be treated with gore® excluder® endoprosthesis.It was stated that the aneurysm was not thrombosed and that the maximum diameter of the aneurysm was 75mm.After the trunk-ipsilateral leg component was positioned and deployed without issue, it was stated that the physician experienced difficulty when advancing the guidewire into the contralateral gate of the trunk-ipsilateral leg component.The device was reported to have been reconstrained and rotated in order to reorient the contralateral gate into a more optimal position.The physician elected to fully deploy the trunk-ipsilateral leg component and brought a snare catheter up and over the aortic bifurcation with the intention of pulling the guidewire into the gate (coded (b)(4)).It was stated that after the guidewire was snared, the trunk-ipsilateral leg component lost the proximal seal and was pulled into the aneurysm sac.The physician elected to explant the trunk-ipsilateral leg component and converted the patient to an open surgical repair of the vasculature.The patient tolerated the procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected e2 to yes.
|
| |
|