| Catalog Number |
10492730 |
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer stated that repeat testing was performed to confirm correct results which met the clinical picture and a corrected report was issued.The requested data files have been received for investigation.The cause of this event is unknown.
|
| |
|
Event or Problem Description
|
|
The customer reported discrepant sodium results on the rp 500 compared to another siemens lab analyzer.There was no report of injury due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
There is evidence of na+ sensor disturbance from interference associated with exposure to quaternary ammonium compounds (such as benzalkonium or chlorhexadine) starting on (b)(6) 2018 at 9:07 am on rp500 sn (b)(4).Several "sodium interfering substance detected" messages were posted in the events log during this time frame.This also caused na+ sensor calibration failures, question values during this time frame.Per the rp500 operators manual, these substances are known interfering substances to the rp500 na+ sensor and should be avoided.
|
| |
|