| Model Number |
199723740S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
Spinal Column Injury (2081)
|
| Date of Event |
07/06/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that spinal fixation or spondy grade 2 s1-l5 was performed.Six months after surgery the treating physician noticed that one screw (right side) was broken, and the rod had slipped through the screw head in s1 on left side.Right side the screw was broken and had rotated.Also the rod had slipped through the screw head l5 on this side.All locking caps were tightened according to the surgical technique and double checked during surgery for correct torque locking.The rod slippage and breakage has occurred in the months after the surgery.The surgery itself was successful and no postoperative events have occurred that might explain the screw breakage.The patient was treated for a spondy grade 2 at level s1-l5.Primary surgery date (b)(6) 2017.Patient gender female.Dob (b)(6).No other patient data is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Device returned for evaluation.Investigation will be conducted.Follow up will be filed with findings.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Visual examination at the macroscopic level found that the screw shank had fractured.A fracture analysis was conducted on the returned device.It was noted that the fracture analysis exhibits two surface morphologies, a smooth region and a grainy/rough region with progression lines which are indicative of fatigue failure.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.A definitive root cause for the screw shank fracturing cannot be positively determined.However, the fracture analysis report exhibits two surface morphologies, a smooth region and a grainy/rough region with progression lines which are indicative of fatigue failure.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|