| Catalog Number |
222315 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/11/2015
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The complaint device was received and inspected.The sharp tip at the distal end appears to be flattened, confirming this failure.Historically, this type of failure has been observed when the device might have been dropped or device was tapped at an off angle or hit the bone accidentally at an off angle.It cannot be determined at what point in time this failure occurred.A batch record review has been conducted for this lot and the results indicate that this batch of product was processed without incident and therefore there is no internally assignable cause for the reported problem.Further, a review into the mitek complaints system revealed unrelated complaints for this lot of devices that were released to distribution.At this point in time, no corrective action is required and no further action is warranted.Mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.(b)(4).This report is being filed from the remetrex complaint management system as required under mitek's corrective and preventative actions (capa) to file usa fda mdr missed malfunctions.
|
| |
|
Event or Problem Description
|
|
It was reported from the affiliate in (b)(6) that during an unspecified surgical procedure, it was observed that the awl device was bent, broke and the tip buckled after pushing into the bone.It was reported that there was a two minute delay in the surgical procedure and a spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Search Alerts/Recalls
|
|
|