| Catalog Number |
210191 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/09/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
Retractor badly damaged due to saw hitting it.Per the mps, all the retractors were damaged and rigid.Case type: tka.
|
| |
|
Additional Manufacturer Narrative
|
|
Follow-up #1 and final report submitted to update sections based on the results of investigation."reported event: per the mps, all the retractors were damaged.Product not returned.Refer to (b)(4).A review of the device history records indicate 150 devices were manufactured and accepted into final stock on (b)(6) 2017.No non-conformances were identified during inspection.A review of complaints in catsweb and trackwise related to p/n 210191, lot number 37831 shows no additional complaints related to the failure in this investigation.Conclusions: system ready for clinical use.No action is required at this time as there is no indication to suggest a product non-conformity or unanticipated hazard.".
|
| |
|
Event or Problem Description
|
|
Retractor badly damaged due to saw hitting it.Per the mps, all the retractors were damaged and rigid.Case type: tka.
|
| |
|
Search Alerts/Recalls
|