| Model Number |
EGIAADAPTXL |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, when the device was set up during a roux-en-y procedure, a reload was placed onto adapter.The jaws were closed but were unable to open.The device was fired and a high pitch squeeling noise was heard from the handle.The knife blade was retracted but jaws were not able to open.The adapter was taken off with reload attached, and manual tool was used to open the jaws and removed the reload.They opened new device and used without issue.There was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.No visual abnormalities were noted.A pmv representative single use loading unit (sulu) was inserted onto the adapter with a pmv drive handle and battery.The green light signaling the attachment of the adapter illuminated indicating that the adapter was recognized.The reload cycled properly and was applied to test media where it was able to apply the staple line and transect the media without difficulty.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|