Model Number M0068404030
Medical Device Problem Code Misfire (2532)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/12/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exact patient age is unknown; however the patient was reported to be over 18 years old.(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 550 laser fiber was used during a ureteral lithotomy under rigid ureteroscopic procedure in the ureter performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, it was noted that the laser fiber short circuited and sparks were observed near the connector.Additionally, the laser fiber stopped firing at this time.There was no burn injury to the nurse/physician.The procedure completed with another accumax 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
One accumax 550 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared used.Additional examination under magnification revealed that the pattern on the tip face was consistent with a slight degradation.There was no damage along the body of the fiber and there was no evidence of a misfire near the connector.Visual examination of the fiber face within the sma connector under magnification revealed that there was debris on the fiber face and that only a small amount of light was coming through.The condition of the returned device would cause the device to emit insufficient energy transmission during ablation thereby confirming the event.The noted damages indicate difficulty was experienced during the procedure and was likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event or Problem Description
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 550 laser fiber was used during a ureteral lithotomy under rigid ureteroscopic procedure in the ureter performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, it was noted that the laser fiber short circuited and sparks were observed near the connector.Additionally, the laser fiber stopped firing at this time.There was no burn injury to the nurse/physician.The procedure completed with another accumax 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCUMAXTM SINGLE USE HOLMIUM LASER FIBER
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
MDR Report Key7757905
Report Number3005099803-2018-02491
Device Sequence Number13905585
Product Code GEX
UDI-Device Identifier08714729784609
UDI-Public08714729784609
Combination Product (Y/N)N
Initial Reporter CountryCN
PMA/510(K) Number
K082928
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/07/2020
Device Model NumberM0068404030
Device Catalogue Number840-403
Device Lot NumberC000004849
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer09/18/2018
Initial Report FDA Received Date08/07/2018
Supplement Report FDA Received Date10/15/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-