| Model Number |
M0068404030 |
| Medical Device Problem Code |
Misfire (2532)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/12/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exact patient age is unknown; however the patient was reported to be over 18 years old.(b)(4).The device has been received for analysis; however the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 550 laser fiber was used during a ureteral lithotomy under rigid ureteroscopic procedure in the ureter performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, it was noted that the laser fiber short circuited and sparks were observed near the connector.Additionally, the laser fiber stopped firing at this time.There was no burn injury to the nurse/physician.The procedure completed with another accumax 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|
Additional Manufacturer Narrative
|
|
One accumax 550 laser fiber was received for evaluation.Visual examination revealed that the exposed glass fiber tip measured approximately 3.5 mm and appeared used.Additional examination under magnification revealed that the pattern on the tip face was consistent with a slight degradation.There was no damage along the body of the fiber and there was no evidence of a misfire near the connector.Visual examination of the fiber face within the sma connector under magnification revealed that there was debris on the fiber face and that only a small amount of light was coming through.The condition of the returned device would cause the device to emit insufficient energy transmission during ablation thereby confirming the event.The noted damages indicate difficulty was experienced during the procedure and was likely due to anatomical or procedural factors such as tortuous anatomy or maneuvering of the device.Therefore, a review and analysis of all available information indicated that the most probable root cause is operational context.A review of the device history record (dhr) was performed; no anomalies were noted.A search of the complaint database revealed that no similar complaints exist for the specified lot.
|
| |
|
Event or Problem Description
|
|
It was reported to boston scientific corporation on (b)(6) 2018 that an accumax 550 laser fiber was used during a ureteral lithotomy under rigid ureteroscopic procedure in the ureter performed on (b)(6) 2018.According to the complainant, during the procedure and outside the patient, it was noted that the laser fiber short circuited and sparks were observed near the connector.Additionally, the laser fiber stopped firing at this time.There was no burn injury to the nurse/physician.The procedure completed with another accumax 550 laser fiber.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
| |
|