| Model Number |
BI-700-00028-120 |
| Medical Device Problem Code |
Communication or Transmission Problem (2896)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/11/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.Onsite functional and visual examination was performed by a manufacturer representative.The umbilical cable was replaced and the issue was resolved.The system passed a system checkout and was returned to an operational condition.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the mobile view station (mvs) was showing the error "image frame rates below recommended levels, please reconnect cable between [imaging system] and mvs and reboot mvs." several restarts did not fix the issue.Umbilical reported as not grossly deformed but very high use.There was no patient present.
|
| |
|
Additional Manufacturer Narrative
|
|
The cable was returned to the manufacturer.Functional testing found that the component had no issues and performed within specifications.The cable passed all continuity tests.If information is provided in the future, a supplemental report will be issued.
|
| |
|