| Model Number |
2067 |
| Medical Device Problem Codes |
Communication or Transmission Problem (2896); Electrical Shorting (2926)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Analysis summary: analysis confirmed the customer comment that the radiofrequency head was unable to obtain any green lights and the head failed its uplink tests.It was noted that the head's lens was cracked.The head passed functional tests with a known good cable.The head is to remain in service stock for future repair.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the radiofrequency programmer head was intermittently unable to get any green lights and this was attributed to a possible short in its cable.The head was returned for service.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|