Model Number 2067
Medical Device Problem Codes Communication or Transmission Problem (2896); Electrical Shorting (2926)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Analysis summary: analysis confirmed the customer comment that the radiofrequency head was unable to obtain any green lights and the head failed its uplink tests.It was noted that the head's lens was cracked.The head passed functional tests with a known good cable.The head is to remain in service stock for future repair.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the radiofrequency programmer head was intermittently unable to get any green lights and this was attributed to a possible short in its cable.The head was returned for service.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARELINK
Common Device Name
PROGRAMMER, PACEMAKER
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key7758683
Report Number2182208-2018-01436
Device Sequence Number1391985
Product Code KRG
UDI-Device Identifier00721902290121
UDI-Public00721902290121
Combination Product (Y/N)N
Initial Reporter StateIA
Initial Reporter CountryUS
PMA/510(K) Number
P890003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number2067
Device Catalogue Number2067
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/18/2018
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date08/07/2018
Supplement Report FDA Received Date10/03/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-