| Model Number |
A3SR01 |
| Medical Device Problem Code |
Appropriate Device Problem Term/Code Not Available (3191)
|
| Health Effect - Clinical Codes |
Pain (1994); Twiddlers Syndrome (2114)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during use of the implantable pulse generator (ipg) system, the patient complained of pain at the site of the pacemaker.X-ray revealed that the lead slack that was present post implant was wrapped around the ipg in the pocket and the pacemaker positioning had changed, with appearance to have been manipulated in the pocket.Upon request, the ipg system was explanted.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|