Model Number A3SR01
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Pain (1994); Twiddlers Syndrome (2114)
Date of Event 07/13/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during use of the implantable pulse generator (ipg) system, the patient complained of pain at the site of the pacemaker.X-ray revealed that the lead slack that was present post implant was wrapped around the ipg in the pocket and the pacemaker positioning had changed, with appearance to have been manipulated in the pocket.Upon request, the ipg system was explanted.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVISA SR MRI SURESCAN
Common Device Name
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
CH  1131
MDR Report Key7759685
Report Number9614453-2018-02781
Device Sequence Number16473671
Product Code NVZ
UDI-Device Identifier00763000059620
UDI-Public00763000059620
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980035
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2018
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date05/14/2019
Device Model NumberA3SR01
Device Catalogue NumberA3SR01
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/14/2018
Supplement Date Received by Manufacturer08/08/2018
Initial Report FDA Received Date08/07/2018
Supplement Report FDA Received Date10/02/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age18 YR
-
-