| Model Number |
PCB00 |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Positioning Failure (1158)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
04/10/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was not implanted.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection using microscope magnification showed the plunger and pushrod in advanced positions.Lack of viscoelastic material was observed on cartridge.Dent/distortion was observed at the cartridge tip section.Lens was also observed damaged and stuck in cartridge.No assembly error and/or defect were observed in the preloaded device related to manufacturing process.The condition in which the sample returned is consistent with a product that was handled and prepared for a surgical process.The complaint issue reported as uncontrolled delivery could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.Labeling review: the directions for use (dfu) was reviewed.The dfu adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
A pre-loaded iol was evaluated and found there was dent/distortion observed at the cartridge tip section.Lens was also observed damaged and stuck in cartridge.Initially, it was reported that a pcb00 3.5 diopter lens shot out of the insertion system before entering the patient''s operative eye.Reportedly, there was no patient contact, and this occurred during handling and prior to insertion.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa (b)(4).
|
| |
|
Search Alerts/Recalls
|