Model Number 1863
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/16/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
 
Event or Problem Description
The customer reported that the device provided an inaccurate reading; spo2 value too low.No consequences or impact to patient were reported.
 
Additional Manufacturer Narrative
The returned sensor was evaluated.During evaluation the sensor passed all visual and functional testing.The sensor was determined to be functioning as designed.
 
Event or Problem Description
The customer reported that the device provided an inaccurate reading; spo2 value too low.No consequences or impact to patient were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LNCS DC-I
Common Device Name
OXIMETER
Manufacturer (Section D)
MASIMO - 40 PARKER
40 parker
irvine CA 92618 1604
MDR Report Key7760299
Report Number2031172-2018-00421
Device Sequence Number1230753
Product Code DQA
UDI-Device Identifier00843997000406
UDI-Public00843997000406
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K051212
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Biomedical Engineer
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 07/16/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1863
Device Catalogue Number1863
Device Lot Number18A14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2018
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer08/22/2018
Initial Report FDA Received Date08/07/2018
Supplement Report FDA Received Date09/11/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-