Lot Number 4396507
Medical Device Problem Code Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Codes Nausea (1970); Hot Flashes/Flushes (2153); Dizziness (2194)
Type of Reportable Event Serious Injury
Event or Problem Description
On (b)(6) 2018, the lay user/patient contacted lifescan (lfs) usa, alleging that her onetouch ultramini meter displayed inaccurately erratic results.The complaint was classified based on the customer care advocate (cca) documentation.The patient reported that the subject meter started reading inaccurately at an unknown time on (b)(6) 2018; however, she was unable to recall the results.The patient manages her diabetes with janumet 50/1000mg oral medication.She denied making any changes to her usual diabetes management routine as a result of the alleged issue ; she indicated that she stuck to her diet and exercised.She reported that an unknown time after obtaining the alleged inaccurately erratic results, she developed symptoms of ¿nauseous, dizzy, light headed and hot flashes¿.She reported that during a doctor¿s office visit her doctor did ¿nothing¿ as he/she was trying to figure out what was wrong.No other treatment for the symptoms was reported.At the time of troubleshooting, the cca noted that the subject meter was set to the correct unit of measure.The patient did not have control solution to test the meter and test strips.Replacement products, including control solution, were sent to the patient.This complaint is being reported because the patient reportedly developed signs and/or symptoms suggestive of a serious injury adverse event, after obtaining alleged inaccurately erratic blood glucose results on the subject meter.
 
Additional Manufacturer Narrative
The lay user/patients meter and test strips have been returned and evaluated by lifescan product analysis with the following findings: the meter passed all testing with no faults found.The reported issue could not be confirmed.Lifescan also conducted an evaluation of the test strip lot and concluded that this lot did not breach the thresholds set for escalation and no systemic issue was observed.Analysis was not possible for the returned test strips due to unknown storage and handling preventing the allegation being physically investigated.If lifescan obtains additional information regarding this complaint, a follow-up report will be submitted.At this time, lifescan considers this matter closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OT ULTRAMINI METER
Common Device Name
GLUCOSE MONITORING SYS/KIT
Manufacturer (Section D)
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
gubelstrasse 34
zug 6300
SZ  6300
MDR Report Key7760840
Report Number3008382007-2018-02246
Device Sequence Number16583648
Product Code NBW
UDI-Device Identifier00353885008167
UDI-Public00353885008167
Combination Product (Y/N)Y
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K061118
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/02/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Lot Number4396507
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/06/2018
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date08/02/2018
Device Age16 MO
Initial Date Received by Manufacturer 08/02/2018
Supplement Date Received by Manufacturer08/02/2018
Initial Report FDA Received Date08/08/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Life Threatening;
-
-