Catalog Number PLC161200
Medical Device Problem Code Expulsion (2933)
Health Effect - Clinical Code Aneurysm (1708)
Date of Event 06/29/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The review of the manufacturing paperwork verified that this lot met all pre-release specifications.Refer to field for the results of the imaging evaluation.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2016, the patient presented with a ruptured abdominal aortic aneurysm that was treated with gore® excluder® aaa endoprostheses in an emergency procedure.On (b)(6) 2018, a type iii endoleak on the patients left was discovered due to insufficient overlap between two contralateral leg endoprostheses.It was also stated that the aneurysm diameter increased from 82-83mm to 84mm.On (b)(6) 2018, a reintervention was successfully performed by relining the separated components with an additional contralateral leg endoprosthesis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC EXCLUDER AAA ENDOPROSTHESIS (LOW PROFILE)
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
MDR Report Key7760893
Report Number2953161-2018-00070
Device Sequence Number19271821
Product Code MIH
UDI-Device Identifier00733132618705
UDI-Public00733132618705
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
P020004
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2018
Device Catalogue NumberPLC161200
Device Lot Number14471833
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date08/08/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PXC141000/(B)(4)
Outcome Attributed to Adverse Event Required Intervention;
Patient Age60 YR
-
-