Model Number 29525
Medical Device Problem Code Power Problem (3010)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Information (3190)
Date of Event 07/10/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The customer states the unit has physical damage.Upon triage, the following issue was found: a damaged power cord with exposed copper wiring.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
An evaluation of the scd 700 compression system was performed for the reported condition of, ¿the unit has physical damage¿.The unit was triaged and service found that the power cord had exposed copper wires.Upon receipt the unit had a power cord and no other accessories.The unit was able to power up on battery power and pass post with no errors.Visual inspection of the power cord revealed the outer insulation was damaged resulting in exposure of the internal insulated wires.Additionally the insulation was damaged on the black (line) and green (ground) wires, resulting in exposed copper.The insulation has a frayed appearance as if it was run over by a wheel and not as if it was cut with a blade or scissors.Power cords periodically require replacement due to age, usage and user damage.Service noted that the power cord insulation " a frayed appearance as if it was run over by a wheel." based on this observation the potential root cause for this issue is customer misuse.All device history records are reviewed for quality in inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Common Device Name
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7761650
Report Number3006451981-2018-00552
Device Sequence Number11211102
Product Code JOW
UDI-Device Identifier10884521129566
UDI-Public10884521129566
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2013
Device Model Number29525
Device Catalogue Number29525
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Initial Date Received by Manufacturer 07/27/2018
Supplement Date Received by Manufacturer07/27/2018
08/08/2018
Initial Report FDA Received Date08/08/2018
Supplement Report FDA Received Date08/10/2018
10/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-