Product complaint #: (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Visual examination of the mis ti cfx fen poly 6x45 reveals signs of operative use as evidenced by superficial markings, with the inner cap (saddle) being missing and the shank of the screw fell apart from its intended location.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.All complaint trends will be evaluated as a part of the depuy spine monthly complaint review meeting.The root cause of the mis ti cfx fen poly 6x45 falling apart cannot be determined from the sample and the information provided.A potential root cause may be excessive force inadvertently placed on the drive feature of the screw during tightening, potentially at a significant angle, resulting in the saddle falling apart during use.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer and no systemic trends were found, this complaint file will be closed with no further action required.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|