Model Number H7493926024350
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/17/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Weight: (b)(6).Device is combination product.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the mid left anterior descending artery.A 3.50 x 24 synergy ii drug-eluting stent was advanced for treatment.However, the device was unable to cross the lesion.When the device was pulled out from a non-bsc guide catheter, the shaft snapped outside the guide catheter.The device was removed by hand.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set equidistant between the proximal and distal markerbands.The crimped stent od (outer diameter) was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the bumper tip showed no signs of damage.A visual and tactile examination of the hypotube found a hypotube break 525 mm distal from the distal end of the strain relief.A hypotube kink was noted just distal to the break site.This type of damages is consistent with excessive force being applied to the delivery system.A visual and tactile examination of the inner and outer shaft polymer extrusion and mid-shaft section found no issues with the extrusion.The bi-component bond showed no signs of damage or strain.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion is operational context as the product meets the design and manufacture specification but due to anatomical/procedural factors encountered during the procedure, performance was limited.(b)(4).
 
Event or Problem Description
It was reported that shaft break occurred.The target lesion was located in the mid left anterior descending artery.A 3.50 x 24 synergy ii drug-eluting stent was advanced for treatment.However, the device was unable to cross the lesion.When the device was pulled out from a non-bsc guide catheter, the shaft snapped outside the guide catheter.The device was removed by hand.The procedure was completed with a different device.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY II EVEROLIMUS-ELUTING PLATINUM CHROMIUM CORONARY STENT SYSTEM
Common Device Name
BIODEGRADABLE POLYMER DRUG ELUTING CORONARY STENT SYSTEM
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7762946
Report Number2134265-2018-07098
Device Sequence Number1318795
Product Code NIQ
UDI-Device Identifier08714729840244
UDI-Public08714729840244
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
P150003
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/25/2018
Device Model NumberH7493926024350
Device Catalogue Number39260-2435
Device Lot Number0021343739
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2018
Initial Date Received by Manufacturer 07/17/2018
Supplement Date Received by Manufacturer08/10/2018
Initial Report FDA Received Date08/08/2018
Supplement Report FDA Received Date08/28/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-