| Model Number |
H7493926024220 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Obstruction/Occlusion (2422)
|
| Date of Event |
08/10/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device is combination product.Device evaluated by mfr.: it is indicated that the device will not be returned for evaluation.If there is any further relevant information obtained, a supplemental mdr will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that jailing occurred.In (b)(6) 2017, the subject presented with stable angina (ccs classification: 4) was referred for elective cardiac catheterization.Target lesion #1 was located in the second obtuse marginal branch (om2) with 99% stenosis and was 20 mm long with a reference vessel diameter of 2.5 mm.Target lesion #1 was treated with pre-dilatation and placement of 2.25x 24 mm study stent.Following post dilatation, residual stenosis was 0%.Following stent implantation in the proximal om2, jailing of the distal left circumflex artery with 95%-99% stenosis was noted and treated with balloon angioplasty.The subject was discharged on aspirin and clopidogrel.
|
| |
|