| Model Number |
72204922 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
07/18/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that due to fiber damage and unexplainable burning of the universal light guide, the screen colored yellow.There was a delay greater than 30 minutes.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation was performed by smith & nephew and could not confirm the customer complaint for there is fiber damage and unexplainable burning of the universal light guide, the screen colored yellow.A visual inspection was performed and showed no damage to both fiber ends of the light guide.No burning was observed.The orange sheathing shows to have several ink spots.The light guide was functionally tested and functioned as intended.When the light guide was attached to the scope it provided a clear sharp image on the monitor.No yellowing was observed.No manufacturing related defects were observed.
|
| |
|
Search Alerts/Recalls
|