U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 4L INFECTION CONTROL URINE DRAINAGE BAG; DRAIN BAG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 4L INFECTION CONTROL URINE DRAINAGE BAG; DRAIN BAG Back to Search Results
Catalog Number 253509A
Medical Device Problem Code Restricted Flow rate (1248)
Health Effect - Clinical Codes Pain (1994); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event or Problem Description
It was reported that the urine would not drain into the drainage bag.Allegedly the patient experienced a hurting bladder.No medical intervention was reported.
 
Additional Manufacturer Narrative
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use states the following: "bard® infection control large-capacity urinary drainage bag ¿ bacteriostatic collection system ¿ bard® ez-lok® sampling port sterile unless package has been opened or damaged.Single use only.Do not resterilize.For urological use only.Caution: federal (u.S.A.) law restricts this device to sale by or on the order of a physician.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with accepted medical practice and applicable local, state and federal laws and regulations.Directions for use: visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or damaged or if any imperfection or surface deterioration is observed, do not use.1.Remove protective cap from drainage tube catheter adapter and connect drainage tube to catheter.2.Position hanger on bedside rail near the foot of the bed.3.Use sheeting clip to secure drainage tube to sheet.Important: position drainage tube in a straight fashion from catheter to drainage bag.4.To empty bag: a.Remove outlet tube from housing: gently squeeze connector arms and pull tube from housing.B.Release clamp and empty bag.C.After emptying, reclamp outlet tube and slide connector into housing until connector arms engage.Note: if specimen is required, see directions for using urine sample port.5.Periodic observations of this system should be made to ensure the urine is flowing freely.If a standing column of urine is observed, check for correct positioning of bag and then for a physical obstruction.If correct positioning or removal of physical obstruction does not allow free flow, the bag may have to be changed.Directions for using bard® ez-lok® sampling port: bard® ez-lok® sampling port accepts a luer-lock or slip tip syringe.1.Kink drainage tubing a minimum of 3 inches below the sampling port until urine is visible under the access site.2.Swab surface of site with antiseptic wipe.3.Using aseptic technique, position the syringe in the center of the sampling port.The syringe should be held perpendicular to the surface of the sampling port (at approximately 80- 100 degree angle).Press the syringe firmly and twist gently to lock the syringe onto the sampling port.Note: improper penetration technique could cause formation of a drop of urine on the surface of the sampling port.Periodic observation of the sampling port is recommended.4.Aspirate desired volume of urine.5.Unkink tubing and send specimen to laboratory.".
 
Event or Problem Description
It was reported that the urine would not drain into the drainage bag.Allegedly the patient experienced a hurting bladder.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4L INFECTION CONTROL URINE DRAINAGE BAG
Common Device Name
DRAIN BAG
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7763818
Report Number1018233-2018-03482
Device Sequence Number1229006
Product Code KNX
UDI-Device Identifier00801741029578
UDI-Public(01)00801741029578
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Catalogue Number253509A
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/20/2018
Supplement Date Received by Manufacturer08/27/2018
Initial Report FDA Received Date08/08/2018
Supplement Report FDA Received Date09/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-