| Model Number |
482400J |
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The customer reports the enteral feeding pump's display keeps on being black.Patient not involved.No further information available at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation of the kangaroo pump was performed and the customer states ¿the enteral feeding pump's display keeps on being black.¿ the unit was triaged and the customer¿s reported condition was confirmed.It was noted that the pump had a faulty motherboard.A review of the device history record shows that this unit was manufactured in 2013 and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
|
| |
|