| Catalog Number |
100/110/085 |
| Medical Device Problem Code |
Gas/Air Leak (2946)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/01/2018
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(6).
|
| |
|
Event or Problem Description
|
|
It was reported that after 30 minutes use of a tracheal tube, air leakage was observed from the cuff.The faulty tube was exchanged with a new one.The tube was checked in according to the instructions for use.Post-use testing was done on the tube and the leak was confirmed by observing the inflated cuff placed in water.There was no patient injury associated with this incident.
|
| |
|
Additional Manufacturer Narrative
|
|
One endotracheal tube was returned for evaluation.The returned sample was visually inspected and no discrepancies were observed.The returned device sample underwent leak testing by inflating and using a syringe.It was submerged under water, and a leak was observed.The most probable root cause is that the product became damaged after it left the shm facilities.This investigation revealed no intrinsic evidence to suggest a root cause related to manufacturing.
|
| |
|