| Model Number |
8886627761 |
| Medical Device Problem Code |
Unraveled Material (1664)
|
| Health Effect - Clinical Codes |
Hernia (2240); No Code Available (3191)
|
| Date of Event |
05/23/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4) if information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post operatively, the patient had palp caudal wound hernia.The surgical wound was opened and the suture has broken and the knot was in tact.All previous suture materials fresh edges have been removed and re sutured lines and closed the skin.The patient was on absolute bed rest with buster collar and was provided medication.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of three devices.Visual inspection and functional testing confirmed there were no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|
| |
|