| Model Number |
35215 |
| Medical Device Problem Code |
Inflation Problem (1310)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/07/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The sample is reported to be available, but has not yet been received by the manufacturer.A review of the device history record is not possible as no lot number was provided.All information reasonably known as of (b)(6) 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by halyard health represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to halyard health.Halyard health has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).Device not returned.
|
| |
|
Event or Problem Description
|
|
Halyard received a single report that referenced at least three different incidences, which were associated with separate units.This is the third of three reports.Refer to 3011270181-2018-00030 for the first report; refer to 3011270181-2018-00031 for the second report.It was reported that there was deflation of the outer balloon, and condensation appeared to be present in the balloon, but there was no observed leak.
|
| |
|
Additional Manufacturer Narrative
|
|
In initial filing, part of the standard verbage was missing.The complete statement is included below.All information reasonably known as of 10 oct 2018 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
One used device was returned for evaluation.There was no observable damage to the device.The cuff was inflated and deflated without problem.The sample was submerged in water and inflated.The absence of bubbles indicated that there was no leak.The cuff, inflation line, and pilot balloon were all manipulated under water and still no leakage was observed.The complaint cannot be confirmed as reported.Root cause is unknown.All information reasonably known as of 14 mar 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The product involved in the report has been returned and is being processed for evaluation.All information reasonably known as of (b)(6) 2019 has been included in this health authority report.Should additional information be obtained, a follow-up health authority report will be provided.The information provided by avanos medical inc.Represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to avanos medical inc.Avanos medical inc.Has no independent knowledge of the event reported but is relaying the information that was provided by the user facility where the incident occurred.This product incident is documented in the halyard health complaint database and identified as complaint (b)(4).This information is submitted pursuant to 21cfr803, in compliance with the medical device reporting requirement and should not be considered to be an admission that an avanos medical inc.Product is defective or caused serious injury.
|
| |
|
Search Alerts/Recalls
|