• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS COBAS 6000 C (501) MODULE; CLINICAL CHEMISTRY ANALYZER Back to Search Results
Model Number C501
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/24/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This event occurred in (b)(6).
 
Event or Problem Description
The customer stated that they received an erroneous result for one patient sample tested for crep2 creatinine plus ver.2 on a cobas 6000 c (501) module - c501.The customer stated that there were no clots or fibrin in the sample.The sample initially resulted with a crep2 value of 3.49 mg/dl and this value was reported outside of the laboratory.The sample was repeated, resulting as 0.93 mg/dl.No adverse events were alleged to have occurred with the patient.The crep2 reagent lot number was 32825201, with an expiration date of 31-dec-2018.A general reagent issue could be ruled out since calibration and controls are acceptable.
 
Additional Manufacturer Narrative
The field service engineer checked the analyzer wash station levels and these were ok.The customer has not reported any additional issues.
 
Additional Manufacturer Narrative
The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COBAS 6000 C (501) MODULE
Common Device Name
CLINICAL CHEMISTRY ANALYZER
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key7766976
Report Number1823260-2018-02664
Device Sequence Number2061417
Product Code JJE
UDI-Device Identifier04015630919215
UDI-Public04015630919215
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K060373
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberC501
Device Catalogue Number04745914001
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/24/2018
Supplement Date Received by Manufacturer07/24/2018
07/24/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date08/21/2018
10/29/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age30 YR
-
-