| Model Number |
NGP0025 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
A surgery center reported mislabeling on two disposable vitrectomy cutters.A brief description from the surgery center indicated they ordered size 25-gauge disposable vitrectomy cutters but upon opening the package, the actual vitrectomy cutter had size 20-gauge engraved on the device.No indication of patient contact.This is 1 of 2 reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 08/10/2018.Device returned to manufacturer? yes.Device evaluation: the disposable vitrectomy cutter was returned within its original packaging confirming the reported lot number of 60084881.A visual inspection of the returned device showed both pouches were opened; no biological fluids or indications that the item was used were observed.Failure mode of mislabeled sample was confirmed as a 20ga vit cutter was returned along with as 25ga labeling.The reported mislabeling issue was confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
Johnson & johnson surgical vision (jjsv) has voluntarily initiated a voluntary recall because a possibility exists that due to a packaging error, a 20 gauge vitrectomy cutter may be found in a 25 gauge package.Use of a vitrectomy cutter that is a different size than expected could lead to the need to alter the surgical technique, including corneal incision enlargement.Corrective and preventive actions will be implemented as required.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|