| Model Number |
NGP0025 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
07/13/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
A surgery center reported mislabeling on two disposable vitrectomy cutters.A brief description from the surgery center indicated they ordered size 25-gauge disposable vitrectomy cutters but upon opening the package, the actual vitrectomy cutter had size 20-gauge engraved on the device.No indication of patient contact.This is 2 of 2 reports.
|
| |
|
Additional Manufacturer Narrative
|
|
Device available for evaluation? yes.Returned to manufacturer on: 08/10/2018.Device returned to manufacturer? yes.Device evaluation: the disposable vitrectomy cutter was returned within its original packaging confirming the reported lot number of 60084881.A visual inspection of the returned device showed both pouches were opened; no biological fluids or indications that the item was used were observed.Failure mode of mislabeled sample was not confirmed as product a 25ga cutter was returned along with 25ga labeling.The reported mislabeling issue was not confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
|
| |
|