• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC WHITE STAR HIGH SPEED VIT CUTTER; DISPOSABLE VITRECTOMY CUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC WHITE STAR HIGH SPEED VIT CUTTER; DISPOSABLE VITRECTOMY CUTTER Back to Search Results
Model Number NGP0025
Medical Device Problem Code Device Markings/Labelling Problem (2911)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/13/2018
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
A surgery center reported mislabeling on two disposable vitrectomy cutters.A brief description from the surgery center indicated they ordered size 25-gauge disposable vitrectomy cutters but upon opening the package, the actual vitrectomy cutter had size 20-gauge engraved on the device.No indication of patient contact.This is 2 of 2 reports.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 08/10/2018.Device returned to manufacturer? yes.Device evaluation: the disposable vitrectomy cutter was returned within its original packaging confirming the reported lot number of 60084881.A visual inspection of the returned device showed both pouches were opened; no biological fluids or indications that the item was used were observed.Failure mode of mislabeled sample was not confirmed as product a 25ga cutter was returned along with 25ga labeling.The reported mislabeling issue was not confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WHITE STAR HIGH SPEED VIT CUTTER
Common Device Name
DISPOSABLE VITRECTOMY CUTTER
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC
santa ana CA
MDR Report Key7767518
Report Number3006695864-2018-01654
Device Sequence Number2061420
Product Code MLZ
UDI-Device Identifier05050474534117
UDI-Public(01)05050474534117(17)200905(10)60084881
Combination Product (Y/N)Y
PMA/510(K) Number
K081681
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/05/2020
Device Model NumberNGP0025
Device Catalogue NumberNGP0025
Device Lot Number60084881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2018
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 07/16/2018
Supplement Date Received by Manufacturer08/26/2018
10/25/2020
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date09/18/2018
11/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-