• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.0MM CANNULATED SCREW-LONG THREAD 54MM; SCREW, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.0MM CANNULATED SCREW-LONG THREAD 54MM; SCREW, FIXATION, BONE Back to Search Results
Model Number 207.754
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 07/18/2018
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Patient identifier and weight not available for reporting.Device malfunctioned intra-operatively and was not implanted / explanted.Complainant part is not expected to be returned for manufacturer review/investigation.Initial reporter telephone not available for reporting.Without a lot number the device history records review could not be completed.The investigation could not be completed; no conclusion could be drawn, as no product was received.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported patient underwent a procedure to repair a transverse two-part patella fracture on (b)(6) 2018 in which the 4.0mm cannulated screw system.The surgeon had reduced the fracture and inserted two (2) 54mm 4.0mm cannulated screws.Surgeon then attempted to remove one of the screws when it broke in half.A portion of the broken screw remained well inside the patella.Trephines from the screw removal set were used to remove the broken portion in the patella.Another screw was then inserted.Surgery was completed satisfactorily with a delay of approximately 30 minutes.Concomitant device reported: 4.0mm cannulated screw (part number unknown, lot number unknown, quantity 1).This report is for one (1) 4.0mm cannulated screw.This is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.0MM CANNULATED SCREW-LONG THREAD 54MM
Common Device Name
SCREW, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key7767777
Report Number2939274-2018-53252
Device Sequence Number181858
Product Code HWC
UDI-Device Identifier10886982148575
UDI-Public(01)10886982148575
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K963192
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number207.754
Device Catalogue Number207.754
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/18/2018
Supplement Date Received by Manufacturer09/04/2018
Initial Report FDA Received Date08/09/2018
Supplement Report FDA Received Date09/04/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age25 YR
-
-