| Model Number |
E7506- |
| Medical Device Problem Codes |
Difficult to Remove (1528); Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Codes |
Burn(s) (1757); No Known Impact Or Consequence To Patient (2692); Partial thickness (Second Degree) Burn (2694)
|
| Date of Event |
07/24/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, after an atrial fibrillation ablation, a patient found to have grounding pad site burn.It was noted that the integrated precision lab did not use a fluoroscopy.Also noted that the customer was using a flexibility catheter, 3.5 mm irrigated tip.The returned patch was placed 1 inch right of the spine at the apex of scapula.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, after an atrial fibrillation ablation, a patient had found to have grounding pad site second degree burn.Topical ointment was applied on the burn site.The return patch was placed 1 inch right of the spine at the apex of scapula.It was noted that the integrated precision lab did not use a fluoroscopy.Also noted that the customer was using a flexibility catheter, 3.5 mm irrigated tip with intracardiac echocardiography the entire procedure.Generator setting of 30-50 w.No further information available at this time.
|
| |
|